NANOPHOR 05 01

Russia’s first serially produced open DNA sequencer

NANOPHOR 05

An eight-capillary genetic analyzer for Sanger sequencing and DNA fragment analysis with laser-induced fluorescence detection.

  • 8 capillaries
  • 7 detection channels
  • Open system

Developed by SYNTOL together with the Institute for Analytical Instrumentation of the Russian Academy of Sciences. Serially manufactured in Russia since 2015.

Explore

Key specifications

Eight capillaries, seven channels and a 96-well plate

8
capillaries

parallel sample separation

96
wells

0.2 ml plate format

7
channels

fluorescent-dye detection

up to 750
nucleotides

read length at no less than 98.5% accuracy

Two analytical modes

Sanger sequencing and DNA fragment analysis

The analyzer separates fluorescently labelled DNA fragments in a polymer matrix and records the signal in seven detection channels. Results can be exported as .ab1 and .fsa files.

01

Sanger sequencing

DNA Analysis software converts sequencing-reaction peaks into a nucleotide sequence and produces an .ab1 file.

02

Fragment analysis

DNA FA software determines fragment lengths and alleles using size standards, bin panels and allelic ladders, then produces an .fsa file.

Instrument and software

Instrument and software for data acquisition and analysis

Eight capillaries

The analyzer uses a 96-well plate and separates samples in eight capillaries in parallel.

Seven detection channels

The optical system records fluorescent dyes in seven spectral channels.

Four software applications

The software set includes instrument control, DNA Analysis, DNA FA and the PAR²SEQ sequence editor.

Standard data formats

Export to .ab1 and .fsa allows further processing in software that supports these formats.

Operating principle

How the analytical result is produced

  1. 01

    Sample loading

    Prepared reactions are placed in a 96-well plate.

  2. 02

    Electrokinetic injection

    DNA fragments are introduced from the wells into the capillaries by an electric field.

  3. 03

    Separation

    Fragments are separated by length in a polymer matrix under high voltage.

  4. 04

    Fluorescence detection

    A 488 nm laser excites the dyes and the optical system records the signal.

  5. 05

    Data processing

    Software produces a DNA sequence or a fragment profile.

Example data

Results obtained using NANOPHOR 05

STR fragment analysis

A VeriFiler™ Express positive control processed in software for STR-profile analysis.

ITS-region sequencing

The Glycyrrhiza inflata ITS region was sequenced using PDMA-6 polymer and 50 cm capillaries.

Technical specifications

NANOPHOR 05 technical parameters

Separation and detection

Number of capillaries
8
Fluorescence detection channels
7
Laser
solid-state, 488 nm, 110 mW
Voltage range
0.1–15 kV
Thermostat temperature
30–70 °C, accuracy ±0.03 °C

Analytical output

Read length
up to 750 nucleotides at no less than 98.5% accuracy
Fragment analysis up to 460 nucleotides
up to 34 minutes
Maximum fragment length
at least 1,200 nucleotides
Data export
.ab1 and .fsa
Plate format
96 × 0.2 ml

Installation and operation

Dimensions
630 × 600 × 630 mm
Weight
no more than 50 kg
Power consumption
300 W
Computer connection
USB, Windows 10
Operating conditions
15–30 °C, relative humidity 20–80%
Warranty service
2 years

Current method recommendations and the user manual define sample preparation, calibration, reagent and maintenance requirements.

Consultation

Discuss your sequencing or fragment-analysis workflow

The SYNTOL team can help select the instrument configuration, reagents and method documentation for your laboratory.